TLA 700 Series P6417 & P6418 Probe Accessory Instructions TLA 700 Series
P6417 & P6418
Probe Accessory
Instructions
070-9948-01
Leadset Connector Kits
These instructions support several probe accessory
kits, which contain leadset connectors for the
P6417 and P6418 Probes. The leadset connectors
simplify connections between the logic analyzer
and your target system. See Figures 1 and 2 for
examples.
For further information about connecting the
probe to the target system, refer to the TLA 700
online help or to the P6417 & P6418 Logic
Analyzer Probe Instructions.
KlipChip
connector
2-to-1
connector
Eight-channel
leadset with
barrel connectors
Figure 1: Leadset connections (barrel connectors shown)
Figure 2: Single-channel leadset connection
Parts Lists
Kit contents are listed below.
P6417 & P6418 Probe accessories kit (020-2198-XX)
Use these connectors to connect the probes Probe
leadsets to square pins or KlipChip connectors.
2
Eight-channel leadsets (barrel connectors)
1
Single-channel leadset (barrel connector)
20
SMT KlipChip connectors
P6417 & P6418 Probe 34-channel probe interface
kit barrel connectors (020-2199-XX)
Use these connectors to connect the probe leadsets
to square pins or KlipChip connectors.
4
Eight-channel leadsets (barrel connectors)
2
Single-channel leadsets (barrel connectors)
1
Set of twelve 2-to-1 signal/ground leadsets
(barrel connectors)
1
Set of four 4-to-1 signal/ground leadsets
(barrel connectors)
P6417 & P6418 Probe 34-channel probe interface kit
mini-PV connectors (020-3000-XX)
Use these connectors to mass-terminate probe
leadsets into a mini-PV terminal connector. One 2
 20 mini-PV terminal connector can
accommodate three eight-channel leadsets.
The mini-PV terminal connectors are designed for
use with 0.025-inch square pins set on 0.1-inch
centers.
4
Eight-channel leadsets (mini-PV connectors)
2
Single-channel leadsets
(mini-PV connectors)
1
Set of twelve 2-to-1 signal/ground leadsets
(mini-PV connectors)
1
Set of four 4-to-1 signal/ground leadsets
(mini-PV connectors)
1
Set of assorted mini-PV terminal connectors
Safety Summary
Refer to the General Safety Summary in the
TLA 700 Series Logic Analyzer Installation
Manual for applicable safety information. Copyright
©
Tektronix, Inc. Printed in U.S.A.
Wilsonville, OR 97070-1000
Connection Procedure
Use the following connection procedure for either
barrel connectors or mini-PV connectors. Steps
that are specific to only one type of connector are
noted.
To avoid static discharge damage, use a grounding
wrist and foot strap, and follow appropriate
antistatic precautions.
1. Power off the system under test.
2. Connect the logic analyzer probe leadsets to
the accessory leadset connectors. See Figure 3.
CAUTION.

Be sure that the logic analyzer
podlets are connected to the 1
 8 signal
row and not to the ground row. Connecting
signals to the ground row can result in
damage to the logic analyzer or the system
under test.
3. For mini-PV leadset connectors, insert the
leads into the mini-PV terminal connector. See
Figure 4. A properly inserted lead snaps into
place and cannot be withdrawn. An
improperly installed lead slides out easily.
If a lead is inserted in the wrong location, the
lead can be removed by gently lifting the
small plastic tab until the lead can be with-
drawn. The tab can be lifted with a small pick
or pointed knife tip. Do not lift the tab too far
or too often, or the tab will break off or
permanently deform.
4. If necessary, use the 2-to-1 or 4-to-1
connectors to connect multiple leads to a
single pin on the system under test. Figure 1
on page 1 shows an example of a 2-to-1
connector in use.
5. Insert the leadset connector(s) onto the pins on
the system under test.
6. Power on the system under test.
Ground lead
(black connector; terminates
in connector ground bus)
Channel leads
Ground pins
(bottom row)
Signal pins
(top row)
Figure 3: Eight-channel leadset connector
Mini-PV
terminal
connector
Mini-PV lead
(crimped side
outward)
Lead should
snap into place.
Figure 4: Mini-PV leads and terminal connector
Characteristics
The characteristics listed in Table 1 apply to all kit
connectors unless otherwise noted.
Table 1: Characteristics
Recommended
usage
TTL and CMOS levels only.
Ground leads should be connected to
ground of system under test.
Not recommended for signals with
edge rates > 2 V/ns.
Maximum clock
speed
50 MHz
AC loading
< 5 pF per channel as seen by the
device under test (plus podlet)
DC loading
None
Termination
(020-2198-XX
and 020-2199-XX
only)
Each signal lead on the eight-channel
leadset contains a 100
W series
termination near the end of the barrel
connector to minimize signal
reflections.
Dimensions
See Figures 5 through 8.
Figures 5 through 8 show the dimensions of the
leadset connectors. Both top and side views are
shown.
152 mm
(6.0 in)
25 mm
(1.0 in)
26 mm
(1.02 in)
8 mm
(.3 in)
21 mm
(.9 in)
Figure 5: Eight-channel leadset (barrel connectors)
64
mm
(2.5 in)
18 mm
(.7 in)
22 mm
(.9 in)
7 mm
(.3 in)
Figure 6: Eight-channel leadset (mini-PV connectors)
152 mm
(6.0 in)
14 mm
(.6 in)
5 mm
(.2 in)
3 mm
(.1 in)
21 mm
(.9 in)
Figure 7: Single-channel leadset (barrel connectors)
64
mm
(2.5 in)
14 mm
(.6 in)
5 mm
(.2 in)
3 mm
(.1 in)
Figure 8: Single-channel leadset (mini-PV connectors)