AEMC INSTRUMENTS F3N
ed waveforms where
most clamp-on meters fail.
Typical measurement environ-
ments include office equipment,
high intensity discharge lights,
variable speed drives and most
all other electronic equipment.
The Model F3N also measures
Min/Max values and Peak values
to 1000A for Crest Factor
calculations an indication of
harmonic content. Both models
measure Frequency, which is
used to detect 3rd order
harmonics (180Hz in 60Hz
systems), set generator
governor speed, or check output
frequency on adjustable speed
drives. The Models F1N and
F3N are very rugged and
designed to withstand daily
professional use.
®
True RMS AC Current measurements to 700A
®
Frequency measurements to 10kHz
®
Stores Min/Max and Avg for Current or Frequency for
up to 24 hours (Model F3N)
®
Fast Peak measurement (including a Max Peak) for
Crest Factor in Harmonics (Model F3N)
®
Smooth Function for 3 second running averages in
Current and Frequency (Model F3N)
®
Digital display with fast analog bargraph to show
surge currents
®
Display Current and Frequency simultaneously
®
Auto and manual ranging modes
®
Tapered jaws handle two 500MCM cables or
one 1.5" (38mm) cable
®
Audible beep confirms key functions and new
Min/Max values
®
Battery life indication displayed at power-up
®
Sleep mode to conserve battery life
®
IEC 1010, 600V, Cat. III and agency approvals
®
Supplied with soft protective carrying holster
®
Direct Replacement
for Fluke
Models 31 & 33
ORDERING INFORMATION
CATALOG NO.
AC Clamp-On Meter Model F1N (TRMS, 400A, Hz). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cat. #1207.01
AC Clamp-On Meter Model F3N (TRMS, 400A, Hz, Peak, Smooth, Min/Max) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cat. #1207.03
Includes soft protective carrying holster, 9V Alkaline battery and user manual
S P E C I F I C A T I O N S
Model F3N measuring phase current on a 3 phase motor.
Models F1N & F3N in soft
protective carrying holster
MODELS
F1N
F3N
ELECTRICAL
AC CURRENT (TRMS)
Measurement Range
0.3 to 400Arms (700A for Sine wave)
Resolution
40A
0.01A
400A
0.1A
Accuracy
±2% Reading + 20cts; ±2% Reading + 2cts
Smooth Function
3 sec running average
Min/Max & Avg
Records and displays
min, max &
average values
PEAK MEASUREMENTS
Response
2ms
Range
99.9A, 1000A
Crest Factor
5 @ mid range
Resolution
40A
0.2A
400A
2A
>400A
4A
Accuracy
±3% Reading + 3cts
FREQUENCY
Range
0.5 to 10kHz
Resolution
0.1Hz
Accuracy
±0.2% Reading + 3cts
GENERAL
Digital Display
3
3
/
4
digits LCD display, 4000-count for A,
10,000-count for Hz, (max. reading 9999)
Analog Display
Fast 40-segment bargraph
Power Source
9V Alkaline battery
Jaw Opening
1.575" (40mm)
Maximum Conductor Size
ø Cable: One 1.5" (38mm);
Two 0.95" (24mm) or 2 x 500MCM
Bus Bar: 1.96 x 0.19" (50 x 5mm)
Dimensions
9.11 x 3.82 x 1.73" (232 x 97 x 44mm)
Weight
1.1 lbs (550g)
Storage Temperature
40° to 158°F (-40° to 70°C)
Humidity
40 to 75% RH
Material
Case: Gray polycarbonate 920A, UL 94 V2;
Jaws: Red polycarbonate 500R,
10% fiberglass, UL 94 V0
Lens: Crystal polycarbonate 920A, UL 94 VI
SAFETY
Agency Approvals
UL, CSA and GS
Ratings
IEC 1010, 600V, Cat. III, CE Mark
3
9
8
6
5
4
2
1
7
1. Jaw assembly
2. Data Hold button
3. Lever for jaw
opening/closing
4. LCD display with
bargraph
5. On/Off button
6. Range button
7. Frequency (Hz)
8. Peak/Smooth
button (Model F3N)
9. Min/Max button
(Model F3N)