INSTALLA AND OPERATING INSTRUCTIONS

ch engine crawled the Web.
The web site itself may have changed. You can check the current page or check for previous versions at the Internet Archive. Yahoo! is not affiliated with the authors of this page or responsible for its content.
INSTALLATION AND OPERATING INSTRUCTIONS
OWNERS MANUAL
INSTALLATION AND
OPERATING INSTRUCTIONS
MODEL 6200
LOW VOLTAGE CONTROL CHAIR Quality management
system
registered to
iso 9001:2000
iso 13485:2003
MADE IN AMERICA
Reliance® Medical Products 3535 Kings Mills Road Mason, Ohio 45040-2303 1-800-735-0357 www.reliance-medical.com
SERIAL NUMBER
For your future reference, mark the serial number
in the space provided.
IN-6200
3
TABLE OF CONTENTS
IMPORTANT INFORMATION.............................................4
1. INTRODUCTION.........................................................5
TEChNICAL SPECIFICATIONS.........................................6
2. INSTALLATION...........................................................7
2.1. Unpacking.......................................................7
Chair......................................................................7
2.2. Assembly.........................................................7
headrest......................................................................7
Power Receptacle...................................................7
3. OPERATION..............................................................7
3.1. Stop Switch.....................................................7
3.2. Chair Lift........................................................8
Raising & Lowering the Chair Lift from the
Footswitch...............................................................8

3.3. Chair Top........................................................9
Posistioning the Chair Back.....................................9
Raising & Lowering the Chair Top From the
Footswitch...............................................................9
3.4. Revolving/Rotation..........................................9
3.5. Arm Movement................................................9
3.6. Footrest..........................................................9
3.7. headrest.........................................................9
Operation-No. 18, 22 headrest................................9
Adjustment-No. 18, 22 headrest............................10
Operation-No. 10 headrest....................................10
Operation-No. 11, 21 headrest..............................10
Adjustment-No. 11, 21 headrest............................10
5. LOCATING AND REPLACING FUSES..........................11
5.1. Fuse Replacement..........................................11
TROUBLEShOOTING GUIDE..........................................12
CLEANING AND MAINTENANCE ....................................13
PARTS LIST.................................................................14
Model 6200 Chair Assembly..................................15
Chair Top Assembly...............................................16
Chair Base Assembly.............................................20
Lift Assembly.........................................................22
#18, 22 headrest Assembly...................................24
#10 headrest Assembly.........................................26
#11 headrest Assembly.........................................28
#21 headrest Assembly.........................................30
Wiring Diagram - 6200 .........................................32
LIMITED WARRANTy.....................................................35 TRANSPORTATION / STORAGE CONDITIONS
Temperature: Operating: 59 to 104º F
15 to 40º C
Storage: -22 to +140º F
-30 to +60º C
Relative Humidity: Operating: 5 to 95% without
condensation
Storage: 10 to 100%
IMPORTANT
INFORMATION
SAFETy SyMBLOLS
Type B, Applied Part
Dangerous Voltage / Shock Hazard
Fuse Rating Specification
Protective Earth Ground
Alternating Current-AC
Safe Working Load
Attention, Consult Accompanying
Documents
DANGER, WARNING, or CAUTION
The exclamation point within an equilateral
triangle is intended to alert the user to
the presence of important operating and
maintenance (servicing) instructions in this
Installation and Operating Instructions.
NOTE
Amplifies a procedure, practice, or condition.
!
IN-6200
IN-6200
4
5 NOTE- Note indicates points of particular or
additional information.
NOTE - la note indique des remarques
dinformation particulière ou supplémentaire.
1.3 Should your product not perform properly, or if
you have any questions concerning the use and care
of any Reliance® product, contact the Reliance®
Distributor, where you purchased this product or contact
the Technical Service Department, Reliance® Medical
Products, Inc., 3535 Kings Mills Road, Mason, Ohio
45040-2303, or call (800) 735-0358.
NOTE: Always have the model number and
serial number available before contacting
Reliance® or your authorized Reliance®
Distributor.
NOTE: Ayez toujours le numéro et le numéro
de série de type disponibles avant Reliance®
entrant en contact ou votre distributeur
autorisé de Reliance®.
CLASSIFIED By CANADIAN STANDARDS
ASSOCIATION® WITh RESPECT TO ELECTRIC
ShOCK, FIRE AND MEChANICAL hAZARDS
ONLy IN ACCORDANCE WITh UL 60601-1.
According to Clause 5 in IEC 60601-1, sec 6.8.1, this
unit is classified by the following:

The type of protection against electric shock:
EQUIPMENT energized from an external electrical
power source: CLASS I EQUIPMENT

The degree of protection against electric shock:
TYPE B EQUIPMENT

The degree of protection against harmful ingress of
water: ORDINARY DEGREE

The degree of safety of application in the presence
of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR
or WITH OXYGEN OR NITROUS OXIDE: EQUIPMENT
not suitable for use in the presence of a FLAMMABLE
ANAESTHETIC MIXTURE WITH AIR or WITH OXYGEN
OR NITROUS OXIDE

The mode of operation: CONTINUOUS OPERATION
1. INTRODUCTION
1.1. This manual contains information applicable only
to the Reliance® Model 6200 Chair.
1.2. Whenever you see the symbols shown below,
heed their instructions! Always follow safe operating
and maintenance practices.
DANGER-

ThE DANGER SyMBOL
IDENTIFIES
SPECIAL INSTRUCTIONS OR
PROCEDURES WhICh, IF NOT CORRECTLy
FOLLOWED, COULD RESULT IN LOSS OF
LIFE OR PERSONAL INJURy.
DANGER -
LE SyMBOLE DE DANGER
IDENTIFIE LES INSTRUCTIONS SPÉCIALES
OU LES PROCÉDURES QUI, SINON
CORRECTEMENT SUIVI, POURRAIENT AVOIR
COMME CONSÉQUENCE LA PERTE DE VIE
OU DE BLESSURES.
WARNING-

ThE WARNING SyMBOL
IDENTIFIES
SPECIAL INSTRUCTIONS
OR PROCEDURES WhICh, IF NOT
CORRECTLy FOLLOWED, COULD RESULT
IN PERSONAL INJURy.
A V E R T I S S A N T -
L E S y M B O L E
DAVERTISSEMENT IDENTIFIE LES
INSTRUCTIONS SPÉCIALES OU LES
PROCÉDURES QUI, SINON CORRECTEMENT
SUIVI, POURRAIENT AVOIR COMME
CONSÉQUENCE DES BLESSURES.
CAUTION-
This caution symbol identifies
special instructions or procedures which,
if not strictly observed, could result in
damage to or destruction of equipment.

ATTENTION -
ce symbole dattention
identifie les instructions spéciales ou
les procédures aux lesquelles, sinon
strictement observés, pourrait avoir comme
conséquence les dégâts ou la destruction
du matériel.
IN-6200
IN-6200
4
5 WARNING:
TO PREVENT FIRE OR ShOCK
hAZARD, DO NOT EXPOSE ThIS APPLIANCE
TO RAIN OR MOISTURE.
ATTENTION:
POUR PRÉVENIR LE RISQUE
DINCENDIE OU DE ChOC, NEXPOSEZ PAS
CET APPAREIL À LA PLUIE OU À LhUMIDITÉ.
TEChNICAL SPECIFICATIONS
Rating for 120V Low Voltage Chair
This equipment is rated Class I, Type B
Equipment Model
Low Voltage Chair 6200
Volts Hertz Amps
120V 50/60 5.5
Class I Type B
Rating for 230V Low Voltage Chair
This equipment is rated Class I, Type B
Equipment Model
Low Voltage Chair 6200
Volts Hertz Amps
230V 50 3.0
Class I Type B
U.S. Patent No.
3,661,421
3,792,905
5,467,002
IMPORTANT USER WARNINGS
WARNING- EXPLOSION:
ThIS ChAIR MUST
NOT BE USED IN ThE PRESENCE OF FLAM-
MABLE ANESThETICS.
AVERTISSANT - DÉCOMPOSITION
:

CETTE PRÉSIDENCE NE DOIT PAS êTRE
UTILISÉE EN PRÉSENCE DES ANESThÉSIQUES
INFLAMMABLES
POTENTIAL ELECTROMAGNETIC or OThER
INTERFERENCE
This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices
in the vicinity. However, there is no guaran