Datasheet
=black>
Below is a cache of http://www.mintl.com/media/pdfs/7010B-CapBridge-DS-rev3_080213.pdf. It's a snapshot of the page taken as our search engine crawled the Web.
The web site itself may have changed. You can check the current page or check for previous versions at the Internet Archive.
Yahoo! is not affiliated with the authors of this page or responsible for its content.
Automatic Balancing
Current Comparator Technology
3-4 Terminal Measurements
Lead Compensation Circuit
Overall Accuracy 20 ppm
Range Extension Option with 7020H
MODEL INFORMATION
The Model 7010B offers a wider capacitance range and
increased voltage sensitivity. Minimum voltage
sensitivity for full scale accuracy is 100 volts. The
7010B is a microprocessor controlled, metrology based,
high voltage capacitance bridge. Its operation is fully
automatic. A large vacuum florescent display presents
relevant measuring quantities such as capacitance (Cx)
and dissipation factor (Tan
). Easy to use front panel
keyboard menus allow the operator to select the
number of readings for statistical analysis of uncertainty
calculations at the 95% (2s) level. All measured
parameters can be transmitted over the IEEE488
interface for storage to a computer.
The model 7010B is a capacitance bridge with a ratio of
1000:1 making it ideal for both low and high voltage
applications. The automatic self-balancing feature
facilitates the use of the bridge for accurate load loss
measurements of large high voltage inductive loads.
The 7010B may also be used for calibration of precision
standard capacitors, tan-delta measurements and of
precision potential transformers. Overall accuracy is
<20 ppm in magnitude and phase. The technology is
based on the two-stage-current-comparator-principle.
The 7010B has a capacitance ratio up to 1000:1 and
dissipation (loss tangent) of 0 to 10% with a resolution
of 1 ppm. To accommodate capacitance ratios larger
than 1000:1 an additional two stage range extender,
model 7020H, may be added to increase the ratio to
1,000,000:1. All connections are made on the rear of
the instrument. The model 7010B is fully protected
against transients.
The effect of lead and winding impedance on the
measurement accuracy has been reduced by means of
a built-in lead compensation circuit.
DATA SHEET
MODEL 7010B
High Voltage Capacitance Bridge
Applications:
Shunt Reactor Loss Measurements
Power Transformer Measurements
Calibration of Potential Transformers
Calibration of Low Voltage Std. Capacitors
Calibration of High Voltage Dividers
Calibration of High Voltage Power Capacitors
Inductance Measurements
Measurement of Low Loss, High Voltage Power
Loss Tangent Measurements to 10%
Corona Loss Measurements
Insulator and Dielectric Testing
Specifications:
Capacitance Range
Cs: 10 pF to 10,000 pF
Cx: 10 pF to 10,000,000 pF (10 uF)
Capacitive Ratio 1:1 to 1000:1
Ns: 0 to 1.11110 in steps of 0.000001
Nx: 1 to 1000 in steps of 1, 2, 5
Primary Current
10 Amp Maximum
Secondary (Cs) Current Range
1uA to 10mA
Dissipation Factor Range
0 to 10% in steps of 0.000001
Inductance Range
700mH to 700000H (Q factor > 10)
Test Frequencies
50 and 60 Hz
Accuracy
Ratio: ± 20 ppm for all Cx Ratios
Loss Angle
±1% of Reading ±10 ppm
Display
Large Vacuum Florescent
Reading Update
1 Second
Warm Up Time
< 5 Minutes to Full Rated Accuracy
Operating Environment
18 to 34
°
C, 10 to 80% RH
Operating Power
100, 120, 220, 240V - 50/60 Hz
Product Details
Dimensions:
545 x 435 x 355 mm
Weight:
41 kg
Shipping Weight:
50 kg
Warranty
1 Year Parts & Labor
Form MI 67, Rev. 1, Dated 01-11-15 (QAP 19, App. O)