Phase Behavoir Cell

Yahoo! is not affiliated with the authors of this page or responsible for its content.
Phase Behavoir Cell
www.slb.com/welltesting
*Mark of Schlumberger
Other company, product, and service names
are the properties of their respective owners.
Copyright © 2007 Schlumberger. All rights reserved. 07-WT-115
Phase Behavior Cell
Improved phase volume measurements and enhanced sample visibility
The pressure-volume-temperature
(PVT) phase behavior cell, an
integral part of the Schlumberger
Oilphase-DBR* fluid sampling and
analysis service, is a precision-
machined instrument for studying
fluid phase behavior and mea-
suring related vapor and liquid
physical properties.
The design of the PVT cell incor-
porates numerous features that
provide improved phase volume
measurements and enhanced
sample visibility. The dominant
feature is the glass cylinder that
encapsulates the fluid sample.
During operation, the structural
integrity of the glass cylinder is
maintained by exerting an over-
burden pressure to the exterior
surface of the cylinder.
Among the many advantages of
the mercury-free PVT cell is the
reduced exposure of the sample
to metal components. Contact is
limited to the end-cap surfaces,
which can be custom manufac-
tured from specialty alloys. This
allows the cell to be readily and
economically modified to handle
samples containing the most
troublesome components.
ApplicAtions
Fluid phase behavior studies
Related
vapor and liquid physical
property measurements
Critical extraction experiments
Hydrate experiments
Benefits
Full sample visibility
Large thermal mass that provides
exceptional temperature stability
Direct phase volume
measurements that
eliminate indirect pump
volume measurement errors
Volumetric distortion
reduction with confining
overburden pressure
feAtures
Constant volumetric
calibration over full
operating pressure range
Pyrex
®
sample containment
cylinder with overburden fluid
for safer operation to 15,000 psi
Mercury-free operation
Nonreactive wetted parts
Compatibility with all types
of reservoir fluids
Rapid phase equilibrium with
magnetic mixer
Low-volume piston for
accurate determination of
trace liquid volumes
specifications
Length
41.9 cm [16.5 in]
Weight
113 kg [250 lbm]
Max. OD
25.4 cm [10.0 in]
Sample volume
135.6 cm
3
[ 8.3 in
3
]
Working pressure
69 or 103 MPa [10,000 or 15,000 psi]
Working temperature
30 to 200 degC [22 to 392 degF]
Sample wetted materials
Pyrex, stainless steel, Hastelloy
®
C-276
®
, or specified
Fittings
High Pressure Equipment Company valves and fittings
Phase behavior cell.
Phase behavior cell for studying phase behavior and fluid properties.