Medium-Power General-Purpose SPST Relays
c Scanning
The PXI-2568 is able to maximize throughput in automated test
applications by the use of scanning. Scanning improves throughput
by downloading a list of up to 32,000 connections to the switch and
cycling through the list using an event (trigger) without any
interruption from the host processor. Scanning is most efficiently
accomplished by mating the PXI-2568 with an instrument, such as
the NI PXI-4070 6
1
2
-digit FlexDMM, which issues a trigger after
each measurement.
Software
All National Instruments PXI switch modules are shipped with
NI-SWITCH,
an IVI-compliant driver offering complete
functionality for all switch modules. For additional assistance in
configuring, programming, and managing higher-channel-count
switching systems, NI Switch Executive software offers an easy-to-use,
intelligent switch management and visual routing environment.
Medium-Power General-Purpose SPST Relays
31 independent SPST relays
Maximum switch capacity
Up to 150 VDC, 150 VAC
Up to 2 A switching/2 A carry
Latching relays
32,000-step scanlist for
deterministic scanning
Fully software programmable
145 cycles/s
Tight synchronization with
instruments through hardware triggers
Operating Systems
Windows 2000/NT/XP
Recommended Software
LabVIEW
LabWindows
/CVI
Measurement Studio
for Visual C++
NI Switch Executive
Other Compatible Software
Visual Basic
C/C++
Driver Software (included)
NI-SWITCH
Compliance
UL
CE
NI PXI-2568
NEW
NI PXI-2568 ..................................................................778572-68
Includes switch module and NI-SWITCH driver software
Accessories
NI Switch Executive
Development System ................................................778546-01
Deployment Engine ..................................................778548-00
BUY ONLINE!
Visit
ni.com/info
and enter pxi2568.
Ordering Information
Input Characteristics
Maximum switching voltage
Channel-to-ground..................................... 150 V, CAT I
Channel-to-channel ................................... 150 V
Maximum switching/carry current................... 2 A
Simultaneous channels at
maximum switching current (25 C) .......... 31
Maximum switching power
(per channel) .............................................. 60 W, 62.5 VA
Typical DC path resistance
Initial.......................................................... <0.15
End of life .................................................. 1
Thermal EMF .................................................... 12 µV
Bandwidth (-3 dB)............................................. 40 MHz
Dynamic characteristics
Maximum speed............................................... 145 cycles/s
Relay operate time
Typical........................................................ 1 ms
Maximum................................................... 3.4 ms
Expected relay life
Mechanical ................................................ 100,000,000 cycles
Electrical
10 VDC, 100 mADC, resistive............. 2,500,000 cycles
10 VDC, 1 ADC, resistive.................... 1,000,000 cycles
30 VDC, 1 ADC resistive..................... 500,000 cycles
60 VDC, 1 ADC resistive..................... 100,000 cycles
Physical
Relay types ...................................................... electromechanical, latching
Contact material............................................... palladium-ruthenium, gold covered
I/O connectors .................................................. 62-pin D-Sub
Dimensions....................................................... 10 by 17.4 cm (3.9 by 6.9 in.) Single-slot, 3U
Environment
Operating temperature..................................... 0 to 55 C
Storage temperature ........................................ -20 to 70 C
Relative humidity ............................................ 5 to 85% noncondensing
Pollution degree ............................................... 2
Indoor use only
Safety
This product is designed to meet the requirements of the following standards of safety for electrical
equipment for measurement, control, and laboratory use:
IEC 61010-1, EN 61010-1
UL 3111-1, UL 61010B-1
CAN/CSA C22.2 No. 1010.1
CE Compliance
This product meets the essential requirements of applicable European Directives, as amended for
CE marking, as follows:
Low-voltage Directive (safety) ......................... 73/23/EEC
Electromagnetic Compatibility
Directive (EMC) ................................................ 89/336/EEC
Specifications
Medium-Power General-Purpose SPST Relays
2
National Instruments Tel: (800) 433-3488 Fax: (512) 683-9300 info@ni.com ni.com
© 2004 National Instruments Corporation. All rights reserved. Product and company names listed are trademarks or trade names of their respective companies.
NI Services and Support
NI has the services and support to meet your
needs around the globe and through the
application life cycle from planning
and development through deployment
and ongoing maintenance. We offer
services and service levels to meet
customer requirements in research,
design, validation, and manufacturing.
Visit
ni.com/services
.
Training and Certification
NI training is the fastest, most certain route to productivity with our
products. NI training can shorten your learning curve, save
development time, and reduce maintenance costs over the
application life cycle. We schedule instructor-led courses in cities
worldwide, or we can hold a course at your facility. We also offer a
professional certification program that identifies individuals who
have high levels of skill and knowledge on using NI products.
Visit
ni.com/training
.
Professional Services
Our Professional Services Team is comprised of NI applications
engineers, NI Consulting Services, and a worldwide NI Alliance
Partner Program of more than 600 independent consultants and
integrators. Services range
from start-up assistance to
turnkey system integration.
Visit
ni.com/alliance
.
OEM Support
We offer design-in consulting and product integration assistance
if you want to use our products for OEM applications. For
information about special pricing and services for OEM customers,
visit
ni.com/oem
.
Local Sales and Technical Support
In offices worldwide, our staff is local to the country, giving you
access to engineers who speak your language. NI delivers industry-
leading technical support through online knowledge bases, our
applications engineers, and access to 14,000 measurement and
automation professionals within NI Developer Exchange forums.
Find immediate answers to your questions at
ni.com/support
.
We also offer service programs that provide automatic upgrades to
your application development environment and higher levels of
technical support. Visit
ni.com/ssp
.
Hardware Services
NI Factory Installation Services
NI Factory Installation Services (FIS) is the fastest and easiest way to
use your PXI or PXI/SCXI combination systems right out of the
box. Trained NI technicians install the software and hardware and
configure the system to your specifications. NI extends the standard
warranty by one year on hardware components (controllers, chassis,
modules) purchased with FIS. To use FIS, simply configure your
system online with
ni.com/pxiadvisor
.
Calibration Services
NI recognizes the need to maintain properly calibrated devices for
high-accuracy measurements. We provide manual calibration
procedures, services to recalibrate your products, and automated
calibration software specifically designed for use by metrology
laboratories. Visit
ni.com/calibration
.
Repair and Extended Warranty
NI provides complete repair services for our products. Express repair
and advance replacement services are also available. We offer
extended warranties to help you meet project life-cycle requirements.
Visit
ni.com/services
.
National Instruments Tel: (512) 683-0100 Fax: (512) 683-9300 info@ni.com
ni.com
(800) 433-3488